Internal Assemblies — Front Chassis 4: Repair
141
Completing the Repair
To Complete the Processor PCA Replacement:
1 Enter the device’s serial number, see “Entering the Serial Number” on page 138.
2 Install the most recent software in the appropriate language, see “Installing Software” on page 139.
3 Enable options, see “Enabling Options” on page 140.
4 Give the user the README document that contains instructions for downloading the most recent
HeartStart XL+ Instructions for Use from the Philips’ Documentation and Download web site
(
http://www.philips.com/ProductDocs).
5 Affix the appropriate Hardware Version label to the bottom of the device.
The Processor PCA ships with a set of Hardware Version labels.
a Clean the surface with isopropyl alcohol. Allow it to dry.
b Affix the appropriate label found in the Processor PCA kit to the bottom of the device, see
“Primary Label” on page 10.
6 Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.