67
4
Repair
This chapter describes how to repair the HeartStart XL+ defibrillator/monitor. Details are provided on
disassembling the device, removing and replacing subassemblies, and reassembling the device.
These instructions are intended for use only by the service providers who are specifically trained to
service the HeartStart XL+ defibrillator/monitor.
Overview
This chapter is organized into the following sections:
Repair Philosophy . . . . . . . . . . . . . . . . . p. 67
Calling for Service . . . . . . . . . . . . . . . . . p. 68
Key Components . . . . . . . . . . . . . . . . . p. 69
Repair Notes . . . . . . . . . . . . . . . . . . p. 69
External Assemblies . . . . . . . . . . . . . . . . . p. 72
Internal Assemblies — Introduction . . . . . . . . . . . . . p. 87
Internal Assemblies — Rear Chassis . . . . . . . . . . . . . p. 97
Internal Assemblies — Front Chassis . . . . . . . . . . . . . p. 114
Closing the Case . . . . . . . . . . . . . . . . . p. 151
Who Should Perform Repairs
Only qualified technical personnel who have been trained in the safe and proper servicing of the
HeartStart XL+ should open the defibrillator/monitor case, remove and replace components, or make
adjustments. If your medical facility does not have qualified technical personnel, contact the Response
Center or your local Philips representative.
WARNING: HeartStart XL+ service should only be performed by qualified service personnel, in accordance with this
document, the HeartStart XL+ Service Manual.
Repair Philosophy
The repair philosophy of the HeartStart XL+ is subassembly replacement.
Examples of subassemblies are the printer, the Processor PCA, and selected connectors and other items.
Repairs that involve replacing individual components on a PCA are not supported.
CAUTION: Individual component replacement should not be attempted. Component level repair is not supported
due to the extensive use of surface mount technology and the high parts-density on the circuit boards.
Unauthorized component replacement can impair performance of the HeartStart XL+ and void the
warranty.