EasyManua.ls Logo

Philips HeartStart XL+ - Therapy Switch

Philips HeartStart XL+
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Internal Assemblies — Front Chassis 4: Repair
121
Therapy Switch
Preparation
1 Remove the Therapy Knob. See Therapy Knob on page 86.
2 Loosen and remove the 9/16-inch nut and washer, see Figure 77.
3 Open and separate the case. See “Opening the Case on page 87.
4 Position the device with the display facing down and the Measurement Module to your left.
5 Pivot the Rear Chassis downward. See “Pivoting Rear Chassis Downward” on page 108.
Removal
1 Unplug the ribbon cable from the Processor PCA.
2 Remove the Therapy switch.
Replacement
1 Position the switch, as shown in Figure 77.
Position the red stripe on the ribbon cable on the edge closest to the Therapy Port.
2 Connect the ribbon cable to the Processor PCA without twisting or kinking the cable.
Ensure the connector is fully seated.
3 Replace the washer and nut. Tighten the nut to 12 inch-lb. (1.4 N m).
4 Replace the Therapy Knob. See “Therapy Knob on page 86.
5 Bring the Rear Chassis to normal position. See “Pivoting Rear Chassis Downward on page 108.
6 Close the case. See “Closing the Case on page 151.
To Complete the Replacement:
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.
Figure 77 Therapy Switch

Table of Contents

Related product manuals