Internal Assemblies — Front Chassis 4: Repair
145
Replacement
1 Replace the chassis and the screws. See “Front Chassis Removal and Replacement” on page 142.
2 Replace the Processor PCA. See “Processor PCA” on page 135.
3 Replace the Rear Chassis. See “Front Chassis and Case Access” on page 134.
4 Replace the Rear Chassis Shelf. See “Rear Chassis Shelf” on page 93.
5 Close the case. See “Closing the Case” on page 151.
To Complete the Replacement:
Run Performance Verification and Safety testing as described in the “Performance Verification”