5: Performance Verification Test and Inspection Matrix
158
NBP
Measurement
(NM)
Take a blood pressure measurement on
yourself or another person. See “NBP
Check on page 167.
If you are able to complete a measurement,
the NBP check passes.
NM:x
Example: NM:p
Synchronized
Cardioversion
(SC)
Using an ECG simulator and defibrillator
analyzer, perform the “Synchronized
Cardioversion Test” on page 170.
If all checks pass, and all data within limits:
Sync markers appear on the display, at the
peak or on the falling side of the QRS
Shock delivered on next QRS
Shock delivered 6 J ± 2 J
If applicable, strip prints with the correct
information on it
Delay between the peak of the QRS and
the delivered shock is <
30 ms
If clinicians use an external monitor as the
ECG source, verify that the external
monitor and the device combination will
deliver a synchronized shock within 60 ms
of the peak of the R-wave.
then Synchronized cardioversion test passes.
SC:x
Example SC:p
Pacing Test
(P)
Using a defibrillator analyzer, perform the
“Pacer Test” on page 169.
(70 ppm) 30 mA
(180 ppm) 140 mA
If all checks pass, and all data within limits:
30mA±5mA
140mA±14mA
then Pacing test passes.
P:x
Example P:p
Defibrillator
Test — AC
Power (DA)
(if AC Power
used in
normal
operation)
Using only AC power and a defibrillator
analyzer, run the “Defibrillator Test (AC
Power at 200 J) on page 168.
If all checks pass, and all data within limits: DA:x
Example DA:p
Energy measured by analyzer:
Delivered into 50-ohm test load 200 ± 20 J
Setting displayed by HeartStart XL+: 200 J
Delivered energy Actual delivered energy ± 10%
then Defibrillator Measurement test passes
Defibrillator
Tes t
Battery Power
(DB)
Using only battery power and a
defibrillator analyzer, run the
“Defibrillator Test (Battery Power at
200 J)” on page 168):
If all checks pass, and all data within limits: DB:x
Example DB:p
Energy measured by analyzer:
Delivered into 50-ohm test load 200 ± 20 J
Setting displayed by HeartStart XL+: 200 J
Delivered energy Actual delivered energy ± 10%
then Defibrillator Measurement test passes.
Table 37 Test and Inspection Matrix (Continued)
Tes t Gro up
Name
Test or Inspection to Perform Expected Test Results Data to Record
x = p (pass)
or f (fail)