Troubleshooting Flowcharts 3: Troubleshooting
45
Figure 32 RFU Indicator: Red X
Insert charged battery or
allow the battery to charge
Run Op Check
Check Tech Alarms,
HW error log
Repair
Run Performance
Verification tests
Troubleshoot device
using tables
RFU displays
Hourglass?
No
Device turns on?
Yes
Place device back in service
Is the Red X
blinking?
Yes
No, solid
See
Figure 31
on page 44
Yes
No
Replace
Therapy PCA
Device
turns on?
No
Call Response
Center
Yes
Does the
device chirp?
No
Yes
Run
Op Check
ECG test
passes?
No
Run ECG test
without ECG
cable
Replace
ECG cable
Yes
Op Check
passes?
Yes
No
ECG test
passes?
Yes
No
Insert both AC
power and a
charged battery
AC light
on front panel
ON?
Yes
No
Red X