4: Repair External Assemblies
82
Replacement
1 If you are replacing the plates only, then install the tray plates into the tray.
a Position the paddle tray up side down and front facing you. See Figure 43.
b Holding the tray plate at an angle, place the plates into the left and right pockets, inserting the
tabs through the holes
, .
c Connect the spade connector from the red wire to the marked RED tab
being careful not to
bend the tab.
d Connect the spade connector from the black wire to the marked BLACK tab
.
e Ensure the connectors are fully seated.
f Replace the two uppermost flat-head M4x12 (T20) screws to attach the new plates to the paddle
tray. Tighten the screws to 10 inch-lb. (1.1 N m).
2 Connect the paddle tray to the device. See Figure 43.
a Connect the spade connector from the red wire to the tab with the red wire on the resistor (
).
b Connect the spade connector from the black wire to the tab with the black wire on the
resistor (
).
c Ensure the connectors are fully seated.
3 Place the paddle tray into position on the device.
Line up the screw holes in the paddle tray and tray plates with the threaded inserts on the device. Be
careful not to pinch the wires under the tray.
4 Replace the four remaining flat-head M4x12 (T20) screws.
Tighten the screws to 10 inch-lb. (1.1 N m).
5 Replace the Handle Assembly, see “Handle Assembly on page 78.
To Complete the Replacement:
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.