EasyManuals Logo

Philips IntelliVue MP5 User Manual

Philips IntelliVue MP5
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
7 ECG, Arrhythmia, ST and QT Monitoring Arrhythmia Alarms
126
You will see that
if both the V-Tach Heart Rate Limit and the V-Tach Run Limit are exceeded, a red V-Tach alarm is
generated
if the ventricular heart rate exceeds the V-Tach Heart Rate Limit but not the V-Tach Run Limit, a
yellow Non-Sustain VT alarm is generated.
Ventricular Heart Rate
Number of Consecutive PVCs (PVC Run)
**Pair
PVCs
PVC =
2
** Non-Sustain VT
PVC Run < 8
HR > 100
***V-Tach
PVC Run
8
HR > 100
** Run PVCs
PVC Run > 2
but12
HR
100
** Vent Rhythm
PVC Run > 12
HR
100
V-Tach Heart
Rate Limit
V-Tach Run Limit
Ventricular Rhythm
Run Limit
1 2 3 4 5 6 7 8 9 10 11 12 14 15
100

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals