7 ECG, Arrhythmia, ST and QT Monitoring Arrhythmia Alarms
126
You will see that
if both the V-Tach Heart Rate Limit and the V-Tach Run Limit are exceeded, a red V-Tach alarm is
generated
if the ventricular heart rate exceeds the V-Tach Heart Rate Limit but not the V-Tach Run Limit, a
yellow Non-Sustain VT alarm is generated.
Ventricular Heart Rate
Number of Consecutive PVCs (PVC Run)
**Pair
PVCs
PVC =
2
** Non-Sustain VT
PVC Run < 8
HR > 100
***V-Tach
PVC Run
8
HR > 100
** Run PVCs
PVC Run > 2
but12
HR
100
** Vent Rhythm
PVC Run > 12
HR
100
V-Tach Heart
Rate Limit
V-Tach Run Limit
Ventricular Rhythm
Run Limit
1 2 3 4 5 6 7 8 9 10 11 12 14 15
100