Audible Alarm Indicators 4Alarms
41
If only patient alarms are present, and no INOPs, the patient alarms will use both left and right
sections to flash (for red and yellow alarms) or light for approximately six seconds (for short yellow
alarms). If only INOPs are present, and no patient alarms, red and yellow INOPs will use both left and
right sections to flash but light blue INOPs will always light continuously in the left section only.
Nurse call systems: If configured to do so, red, yellow and light blue alarms are indicated on any
device connected to the nurse call relay.
Audible Alarm Indicators
The audible alarm indicators configured for your monitor depend on which alarm standard applies in
your hospital. Audible alarm indicator patterns are repeated until you acknowledge the alarm by
switching it off or pausing it, or until the alarm condition ceases (if audible alarm indication is set to
non-latching).
WARNING Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarm
volume to a low level or off during patient monitoring may result in patient danger. Remember that
the most reliable method of patient monitoring combines close personal surveillance with correct
operation of monitoring equipment.
No alarms are available on the MP5 when connected to a host monitor (Companion Mode is
indicated). Alarms become active again as soon as the MP5 is disconnected from the host monitor.
When connected to a host monitor (Companion Mode is indicated), no alarm tones are available
on the MP5.
Alarm Tone Configuration
The audible alarm indicators of your monitor are configurable. In the monitor’s Configuration Mode,
you can:
increase the alarm volume of unacknowledged alarms at regular intervals
change the interval between alarm sounds (ISO/IEC Standard alarms only)
change the base volume of the red and yellow alarm tones and the INOP tones
change the alarm sound to suit the different alarm standards valid in different countries.
Traditional Audible Alarms (HP/Agilent/Philips/Carenet)
Red alarms and red INOPs: A high pitched sound is repeated once a second.
Two-star yellow alarms and yellow INOPs: A lower pitched sound is repeated every two seconds.
INOP Lamp
Color
Modulation (how long the lamp is on or off while flashing)
On Off
Yellow 1.0 seconds 1.0 seconds
Red 0.25 seconds 0.25 seconds

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals