Measurement Specifications 31 Installation and Specifications
317
Te m p
Complies with EN 12470-4:2000. Specified without transducer.
Predictive Temperature
CO
2
The CO
2
measurement complies with EN ISO 21647:2004 + Cor.1:2005 (except alarm system; alarm
system complies with IEC 60601-2-49:2001).
Tachycardia Difference to high limit 0 to
50 bpm
5 bpm steps max.
14 seconds
Clamping at 150 to
300 bpm
5 bpm steps
Bradycardia Difference to low limit 0 to
50 bpm
5 bpm steps max.
14 seconds
Clamping at 25 to 100 bpm
5 bpm steps
Invasive Pressure Alarm Specifications Range Adjustment Delay
Temp Performance Specifications
Te m p Range –1 to 45C (30 to 113F)
Resolution
0.1C (0.2F)
Accuracy
±0.1C (±0.2F)
Average Time Constant Less than 10 seconds
Temp Alarm Specifications Range Adjustment
Temp H i gh/ L ow A larms –1 to 45C (30 to 113F) -1 to 35C (30 to 95F), 0.5C (1.0F) steps
35 to 45C (95 to 113F), 0.1C (0.2F) steps
Performance Specifications
Te c hn ol og y Welch Allyn® SureTemp Plus®
Probe Types oral/axillary, rectal
Temperature Measurement Range 26.7 to 43.3C (80 to 110F)
Resolution ±0.1C (±0.2F)
Accuracy
(Oral, axillary and rectal measurements)
±0.1C (±0.2F) (in continuous mode, complies with ASTM 1112-
00