31 Installation and Specifications Measurement Specifications
320
Interfering Gas and Vapor Effects On CO
2
Measurement Values
The specified deviations are valid when the appropriate corrections are switched on and set correctly.
imCO
2
High 2 to 20 mmHg
(0.3 to 3.0 kPa)
steps of 1 mmHg
(0.1 kPa)
less than 14 seconds
awRR High Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
under 20 rpm: 1 rpm
steps
over 20 rpm:5 rpm
steps
less than 14 seconds
awRR Low Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
settings <20 rpm: less than 4 seconds
settings >20 rpm: less than 14 seconds
Apnea delay 10 to 40 seconds 5 second steps set apnea delay time +4 seconds
CO
2
Alarm
Specifications Range Adjustment Delay
Gas or Vapor Gas Level
(% volume fraction)
Additional deviation due to gas interference, measured
at 0 - 40 mmHg CO
2
Nitrous Oxide 60 ±1 mmHg
Halothane 4 ±2 mmHg
Enflurane 5 ±2 mmHg
Isoflurane 5 ±2 mmHg
Sevoflurane 5 ±2 mmHg
Xenon 80 -5 mmHg
Helium 50 ±1 mmHg
Metered dose inhaler
propellants
- not specified for use
Desflurane 15 +5 mmHg
Ethanol 0.1 ±1 mmHg
Isopropanol 0.1 ±1 mmHg
Acetone 0.1 ±1 mmHg
Methane 1.0 ±1 mmHg

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals