153
10
10Monitoring SpO
2
Philips pulse oximetry uses a motion-tolerant signal processing algorithm, based on Fourier artefact
suppression technology (FAST). It provides four measurements:
Oxygen saturation of arterial blood (SpO
2
) - percentage of oxygenated hemoglobin in relation to the
sum of oxyhemoglobin and deoxyhemoglobin (functional arterial oxygen saturation).
Pleth waveform - visual indication of patient’s pulse.
Pulse rate (derived from pleth wave) - detected pulsations per minute.
Perfusion indicator - numerical value for the pulsatile portion of the measured signal caused by
arterial pulsation.
The monitors are also compatible with SpO
2
technologies from other manufacturers. Please refer to the
instructions for use provided with these devices for further information.
SpO
2
Sensors
Familiarize yourself with the instructions for use supplied with your sensor before using it. In
particular, check that the sensor being used is appropriate for your patient category and application
site.
CAUTION Do not use OxiCliq disposable sensors in a high humidity environment, such as in neonatal incubators
or in the presence of fluids, which may contaminate sensor and electrical connections causing
unreliable or intermittent measurements. Do not use disposable sensors on patients who have allergic
reactions to the adhesive.
Applying the Sensor
1 Follow the SpO
2
sensor’s instructions for use, adhering to all warnings and cautions.
2 Remove colored nail polish from the application site.
3 Apply the sensor to the patient. The application site should match the sensor size so that the sensor
can neither fall off, nor apply excessive pressure. When using the M1195A Infant Finger Sensor,
select a finger or toe with a diameter of between 7 and 8 mm (0.27” and 0.31”). When applying a
M1193A neonatal sensor do not overtighten the strap.

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals