Safety and Performance Tests 31 Installation and Specifications
323
Recommended Separation Distance
WARNING The monitor, equipped with a wireless network interface, intentionally receives RF electromagnetic
energy for the purpose of its operation. Therefore, other equipment may cause interference, even if that
other equipment complies with CISPR emission requirements.
In the following table, P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation distance in metres
(m). The values given in brackets are for respiration.
Portable and mobile RF communications equipment should be used no closer to any part of the
monitor, including cables, than the recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with this symbol:
Immunity test
IEC 60601-1-2
test level Compliance level Electromagnetic environment guidance
Conducted RF
IEC 61000-4-6
3V
RMS
150 kHz to
80 MHz
3V
RMS
(1 V
RMS
for
respiration)
Recommended separation distance:
:
for respiration:
:
Radiated RF
IEC 61000-4-3
3V/m
80 MHz to
2.5 GHz
3V/m
(1 V/m for respiration)
For short range radio,
see note
1
.
1.If ECG/SpO
2
signals are acquired from a telemetry device via short range radio the compliance level is 3V/m except in
the range 2.0 to 2.3 GHz where it is 1 V/m.
Recommended separation distance:
80 MHz to 800 MHz
80 MHz to 800 MHz for respiration
:
800 MHz to 2,5 GHz
800 MHz to 2,5 GHz for respiration
2.0 to 2,3 GHz for short range radio
d1,2P=
d3,5P=
d1,2P=
d3,5P=
d2,3P=
d7,0P=
d7,0P=

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals