EasyManuals Logo

Philips IntelliVue MP5 User Manual

Philips IntelliVue MP5
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
185
15
15Monitoring Carbon Dioxide
Use the CO
2
measurement to monitor the patient’s respiratory status and to control patient
ventilation.
There are two methods for measuring carbon dioxide in the patient’s airway:
Mainstream measurement uses a CO
2
sensor attached to an airway adapter directly inserted into the
patients breathing system.
Sidestream measurement takes a sample of the respiratory gas with a constant sample flow from the
patient’s airway and analyzes it with a remote CO
2
sensor. Two variations of this method, with
different sensors, are available as options on the monitor. They are referred to here as sidestream
CO
2
and Microstream CO
2
.
In both cases, the measurement principle is infrared transmission, where the intensity of infrared light
passing the respiratory gas is measured with a photo detector. As some of the infrared light is absorbed
by the CO
2
molecules, the amount of light passing the gas probe depends on the concentration of the
measured CO
2
.
When using a wet ventilator circuit, monitor mainstream CO
2
if available, in preference to sidestream
CO
2
.
The partial pressure is calculated from the gas concentration by multiplying the concentration value
with the ambient pressure.
The measurement provides:
•a CO
2
waveform.
an end tidal CO
2
(etCO
2
) value: the CO
2
value measured at the end of the expiration phase.
an inspired minimum CO
2
(imCO
2
): the smallest value measured during inspiration.
an airway respiration rate (awRR): the number of breaths per minute, calculated from the CO
2
waveform.
Depending on the Max Hold setting configured for your monitor, the etCO
2
numeric shows either
the highest CO
2
value measured within the configured time period (Max Hold set to 10 sec or
20 sec) or the etCO
2
numeric shows breath-to-breath value (Max Hold set to Off).
WARNING Correlation: The etCO
2
readings do not always correlate closely with paCO
2
, especially in neonatal
patients and patients with pulmonary disease, pulmonary embolism or inappropriate ventilation.
Pharmaceuticals in aerosols: Do not measure CO
2
in the presence of pharmaceuticals in aerosols.

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals