18 Trends Screen Trends
218
Auto Horizon to set the horizon for the selected horizon trend to the currently-measured
value
Auto All to reset the horizon for all horizon screen trends to the currently-measured values
Set Horizon to set the horizon to a specific value from a pop-up list.
Setting the Horizon Trend Scale
The horizon trend scale is defined with respect to the horizon. If your horizon is set to 100 and you
select 20 as the horizon scale delta, the upper limit of the horizon scale will be 120 and lower limit 80.
To set the horizon trend scale delta,
1 Select the horizon trend.
2 Select Set Scale Delta and select a value to define the distance between the horizon and the
upper and lower scale limits.
Be aware that changing the horizon trend scale can change the angle of the trend indicator, without the
patient’s condition having changed.
If a measurement exceeds the outer limits of the scale, the wave will be clipped and you must either
reset the horizon or the horizon trend scale to display the values outside the scale limits.
Setting the Time Period for the Trend Indicator Arrow
The time period for which the trend indicator arrow can be set in the Trends window.
1 Select Main Setup > Trends
2 Select HorizonArrowTime.
3 Select 10, 5 or 2 minutes.

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals