Technical Alarm Messages (INOPs) 5 Patient Alarms and INOPs
61
Technical Alarm Messages (INOPs)
If an INOP interrupts monitoring and alarm detection, the measurement numeric will be replaced
by -?-. If an INOP may lead to unreliable measurement values, a ? appears next to the numeric.
The measurement labels and abbreviations for pressure, temperature and SpO
2
INOP messages are
explained in the individual chapters.
INOP Message, Indication Source What to do
ABP INOPS PRESS See <Pressure label> INOPS (under Pressure).
ALL ECG ALARMS OFF ECG/
Arrhythmia
All ECG alarms have been switched off, or the HR alarm
source is not ECG. To resume ECG alarm generation, switch
ECG alarms on or select ECG as the alarms source.
Ao INOPS PRESS See <Pressure label> INOPS (under Pressure).
ART INOPS PRESS See <Pressure label> INOPS (under Pressure).
AWF CHANGE SCALE Spirometry Airway flow signal exceeds range of selected scale. Adjust scale
to display complete wave.
AWP CHANGE SCALE Spirometry Airway pressure signal exceeds range of selected scale. Adjust
scale to display complete wave
AWV CHANGE SCALE Spirometry Airway volume signal exceeds range of selected scale. Adjust
scale to display complete wave.
Bad Serverlink
INOP tone
Monitor 1) An MMS with an incompatible software revision is
connected to the monitor. This combination does not allow
monitoring, OR
2) You cannot use this combination of monitor, MMS and
cable. Switch off the monitor and contact your service
personnel.
BAP INOPS PRESS See <Pressure label> INOPS (under Pressure).
BATT EMPTY
INOP tone, battery LED flashes
During this INOP, alarms cannot be
paused or switched off.
Battery The estimated remaining battery-powered operating time is less
than 10 minutes. Replace the battery immediately.
If the condition persists and the monitor is not connected to
mains power, this INOP is re-issued two minutes after you
acknowledge it.
BATT INCOMPAT
INOP tone
Battery The battery cannot be used with this monitor. Replace with the
correct battery
(M4607A).
BATT LOW
INOP tone
Battery The estimated battery-powered operating time remaining is less
than 20 minutes.
BATT MALFUNCTION
INOP tone, battery LED flashes
During this INOP, alarms cannot be
paused or switched off unless the
monitor is connected to mains power.
Battery The monitor cannot determine the battery status. If this INOP
persists, replace the faulty battery. If the condition persists and
the monitor is not connected to mains power, this INOP is re-
issued two minutes after you acknowledge it.
Place the battery in a different monitor or in a battery charger.
If the same INOP is shown, contact your service personnel.
BATTERIES EMPTY or BATT 1/
BATT 2 EMPTY
INOP tone, battery LED flashes
During this INOP, alarms cannot be
paused or switched off.
Batteries The estimated remaining battery-powered operating time of
the indicated battery or batteries is less than 10 minutes.
Replace the batteries immediately.
If the condition persists and the monitor is not connected to
mains power, this INOP is re-issued two minutes after you
acknowledge it.

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals