Pulse Default Settings 32 Default Settings Appendix
337
Pulse Default Settings
Pulse alarms use the settings of the currently selected Pulse alarm source.
Respiration Default Settings
Pulse Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Alarms Source Auto Auto
Pulse (Label) On
System Pulse SpO
2
Auto
Alarms Off Enabled
Alarm Source Sel. Enabled
Pulse Alarm Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Pulse (SpO2) On
Pulse Alarms Off
High Limit 120 bpm 160 bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
Extr Brady 20 bpm
Brady Clamp 40 bpm 40 bpm 50 bpm 40 60 80
Extr Tachy 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm
Respiration Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
High Limit 30 rpm 100 rpm
Low Limit 8 rpm 30 rpm
Apnea Time 20 sec
Alarms On
Resp On Off
Detection Auto (Trigger Mode)
Respiratory Speed 6.25 mm/s
Color Yellow White