32 Default Settings Appendix ECG, Arrhythmia, ST and QT Default Settings
336
ST Analysis On Off
ST-Index On
ISO Point -80 ms
J Point 48 ms
ST Point J+60
Lead I, II, III, V, aVR, aVL, aVF, V1-6,
MCL Settings
Factory Adult Factory Pedi Factory Neo
ST(Label) On Off
For Alarm Mode = Single-ST
ST
(Label) High +2.0 mm
ST
(Label) Low -2.0 mm
For Alarm Mode = Multi-ST
ST(Label) High +1.0 mm
ST
(Label) Low -1.0 mm
QT Settings Factory Adult Factory Pedi Factory Neo
QT Lead All
QTc High Limit 500 ms 480 ms 460 ms
QTc Limit 60 ms
QTc High Alarm On
QTc High Alarm On
QT Analysis Off
QTc Formula Bazett
Lead-independent ST Settings Factory Adult Factory Pedi Factory Neo