Enabling Automatic Mode and Setting Repetition Time 11 Monitoring NBP
165
CAUTION Use clinical judgement to decide whether to perform repeated series of STAT measurements because of
the risk of purpura, ischemia and neuropathy in the limb with the cuff.
Enabling Automatic Mode and Setting Repetition
Time
1 In the Setup NBP menu, select Mode and select Auto from the pop-up menu.
2 For an automatic measurement, select Repeat Time and set the time interval between two
measurements.
Enabling Sequence Mode and Setting Up The
Sequence
1 In the Setup NBP menu, select Mode and select Sequence from the pop-up menu.
2 Select Setup Sequence to open the Setup Sequence window.
Up to four measurement cycles can be setup which will run consecutively. For each cycle you can
set the number of measurements and the interval between them. If you want to run less than four
cycles in a sequence, you can set the number of measurements for one or more cycles to Off.
3 Select each sequence in turn and select the number of measurements and the time interval between
the measurements.
4 To have measurements continue after the sequence, set the number of measurements for your last
cycle to Continuous and this cycle will run indefinitely.
Stop current STAT measurement and end series
Start/Stop
NBP STAT
Stop Auto, Manual or STAT measurement AND
series
Stop All
Action to be performed NBP Setup menu SmartKeys
Start/
Stop
NBP
STAT
Stop
NBP
Stop
All