14 Monitoring Invasive Pressure Choosing the Pressure Alarm Source
180
Select and set the High Limit and Low Limit for the pressure(s) you have selected.
Extreme Alarm Limits
The extreme pressure alarms, Extreme High and Extreme Low, can be made available for your monitor
in Configuration Mode and are additional to the standard High and Low limit alarms. They are
generated by the active pressure alarm source, and are setup in Configuration Mode by adding a set
value (the value) to the high and low alarm limits. This value can be set for each pressure label
individually.
You need to know which values have been configured for your monitor. Changing the high and low
alarm limits automatically changes the extreme alarm limits within the allowed range.
To see the extreme pressure alarms set for your monitor, in the Setup <Press> menu, see the
menu items
Extreme High and Extreme Low.
The extreme pressure alarms are high priority, red alarms, marked *** in the alarm message.
Sys & Dia systolic and diastolic in parallel
Dia & Mean diastolic and mean in parallel
Sys & Mean systolic and mean in parallel
Sys&Dia&Mean all three pressures in parallel
Menu option Pressure value monitored
Low
Limit
High
Limit
Extreme
Low Limit
Extreme
High Limit
Extreme Low Extreme High

Table of Contents

Other manuals for Philips IntelliVue MP5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP5 and is the answer not in the manual?

Philips IntelliVue MP5 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP5
CategoryMedical Equipment
LanguageEnglish

Related product manuals