32 Default Settings Appendix Predictive Temperature Default Settings
340
Predictive Temperature Default Settings
Invasive Pressure Default Settings
pTemp Settings
1
1.The settings for Prompt Tones, Unit and Color are label dependent
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Preferred Label pToral
Prompt Tones on
Value Lifetime 1 hour
Unit C
Color Yellow
ABP, ART, Ao, BAP,
FAP, P, P1, P2, P3,
P4, UAP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Sys.
High Limit 160/90 (110) 120/70 (90) 90/60 (70) 180/90 (110)
Low Limit 90/50 (70) 70/40 (50) 55/20 (36) 70/50 (70)
Alarms On
Extreme Alarms Disabled
Extreme High 15
10 5
Extreme Low 15
10 5
High Clamp 190/100 (125) 140/80 (100) 105/75 (75)
Low Clamp 80/45 (65) 60/35 (45) 45/15 (30) 65/45 (65)
Scale 150 100 100
Speed 25 mm/s
Mean Only No
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Red
C V P, R A P, L A P,
UVP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On