Invasive Pressure Default Settings 32 Default Settings Appendix
341
Extreme Alarms Enabled
Extreme High 5 5 5
Extreme Low 5 5 5
High Clamp 20/10 (15) 15/5 (10) 15/5 (10)
Low Clamp 0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Scale 30
Speed 25 mm/s
Mean Only Yes
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color light blue (cyan) Blue
PAP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Dia.
High Limit 34/16 (20) 60/4 (26) 60/4 (26)
Low Limit 10/0 (0) 24/-4 (12) 24/-4 (12)
Alarms On
Extreme Alarms Enabled
Enabled Enabled
Extreme High 5 5 5
Extreme Low 5 5 5
High Clamp 45/20 (25) 65/5 (35) 65/5 (35)
Low Clamp 5/-5 (-5) 15/-5 (5) 15/-5 (5)
Scale 30
Speed 25 mm/s
Mean Only No
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Yellow
C V P, R A P, L A P,
UVP Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo