32 Default Settings Appendix CO
2
Default Settings
342
CO
2
Default Settings
ICP, IC1, IC2
Settings
Factory Defaults
Adult Pedi Neo
Alarms from Mean
High Limit 14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On
Extreme Alarms Enabled
Enabled Enabled
Extreme High 10 10 10
Extreme Low 10 10 10
Low Clamp 20/10 (-5) 15/5 (-5) 15/5 (-5)
Mean High Clamp 0/-5 (0) 0/-5 (0) 0/-5 (0)
Scale 30
Speed 25 mm/s
Mean Only Yes
Filter 12 Hz
Mercury Cal. Yes
Artifact Suppr. 60 sec
Unit mmHg
Color Magenta
CO
2
Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
etCO
2
low 30 25
etCO
2
high 50 60
imCO
2
high 4
CO
2
Alarms on
Unit mmHg
Scale 40 mmHg 50
ImCO
2
on Off
Oxygen Corr 16%
Gas Corr Off
Humidity Corr BTPS
Max Hold Off
AwRR On
AwRR Alarms On