Calculating Pulse Pressure Variation 14 Monitoring Invasive Pressure
183
The clinical value of the derived PPV information must be determined by a physician. According to
recent scientific literature, the clinical relevance of PPV information is restricted to sedated patients
receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia.
PPV calculation may lead to inaccurate values in the following situations:
at respiration rates below 8 rpm
during ventilation with tidal volumes lower than 8 ml/kg
for patients with acute right ventricular dysfunction (“cor pulmonale”).
The PPV measurement has been validated only for adult patients
To select an arterial pressure as PPV source:
1 In the Main Setup menu, select Measurements.
2 In the Setup PPV menu, select ABP, ART, Ao, BAP, P or FAP as the arterial pressure source.