Monitor Performance Specifications 31 Installation and Specifications
305
Battery Specifications Operating Time
(with new, fully
charged battery)
Basic monitoring configuration: >4 hours
(Brightness set to Optimum, ECG/Resp, SpO
2
measurements in use, NBP measurement every 15 minutes)
Extended monitoring configuration: >3 hours
(Brightness set to Optimum, ECG/Resp, SpO
2
, Press/
Temp, C O
2
measurements in use, NBP every 15 minutes,
Recorder)
Charge Time
When monitor is off: 4 hours
When monitor is in use: 5 hours and above, depending on
monitor configuration (in some configurations the battery
may not completely recharge in the monitor, in this case the
M8043A Smart Battery Charger should be used)
Indicators Alarms Off red (crossed-out alarm symbol) LED
Alarms
red/yellow/light blue (cyan) LED
On/Standby/Error
green/red LED
AC Power
green LED
Battery LED
red/yellow/green LED
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO
2
modulation tone
4 different alarm sounds
Remote tone for alarms on other beds in network
Tone for Timer expired
Trends Resolution 12, 16, 24 or 32 numerics @ 12 sec, 1 minute, 5 minute
resolution
Information
Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
neonatal extended 12 numerics, 24 hours @ 12 secs or
32 numerics 32 hours @ 1 minute
intensive care extended: 16 numerics 120 hours @ 5 minutes
anesthesia extended 32 numerics 9 hours @ 12 seconds
High-Res Trend Waves Measurements
OxyCRG
HR, SpO
2
, Resp
Resolution
Measurement samples are taken at a resolution of four samples
per second
Update speed
waves are drawn at a speed of 3 cm/minute
Events Information trigger condition and time, event classification and associated
detailed view of episode data
Episode data
4 minutes of high resolution trend
Alarm signal System delay less than 3 seconds
Pause duration
1,2,3 minutes or infinite, depending on configuration
Extended alarm pause
5 or 10 minutes
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity
300 items
Performance Specifications

Table of Contents

Other manuals for Philips IntelliVue MP5