Technical Alarm Messages (INOPs) 5 Patient Alarms and INOPs
73
MSL Power High Monitor The power consumption of the devices connected to the
Measurement Link (MSL) cable is too high. If this situation
continues, the MSL will be switched off. Contact your service
personnel.
MSL Power Off
INOP tone.
Monitor The power consumption of the devices connected to the
Measurement Link (MSL) cable was too high for too long and
the MSL has been switched off. Contact your service personnel.
MSL Power Overload
INOP tone.
Monitor The power consumption of the devices connected to the
Measurement Link (MSL) cable is much too high or there has
been a short circuit. The MSL has been switched off. Contact
your service personnel.
NBP Deactivated
INOP tone.
NBP The NBP measurement label in the measurement device has
been deactivated by deactivating the label in the Measurement
Selection window. The measurement automatically disappears
from the display. To switch the measurement on again,
reactivate the measurement label in the Measurement Selection
window.
NBP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
NBP Remove the cuff from the patient. The NBP hardware is faulty.
Contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop
All
SmartKey is selected.
NBP INTERRUPTED
Numeric is replaced by -?-
INOP tone.
NBP Check the tubing and cuff for leakages or kinks. Check that
you are using the correct cuff size and placement, and that the
correct patient category is selected. Try restarting the
measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop
All
SmartKey is selected.
This INOP arises when the measurement needed longer than
the maximum time for inflation, deflation or the total
measurement.
NBP MEASURE FAILED
Numeric may be displayed with a -?-
INOP tone.
NBP Check that you are using the correct cuff size and placement,
and that the correct patient category is selected. Try restarting
the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop
All
SmartKey is selected.
Check the condition and suitability of the patient for NBP
monitoring. Use another cuff to continue measuring.
No Central Monit.
INOP tone
Monitor There is a problem with the communication to the network.
Central monitoring is currently not possible (no patient alarms
or information). Check the connection. Contact your service
personnel.
!!/!!! NO DEVICE DATA
NO DEVICE DATA
IntelliBridge Communication with connected device has been lost.
NO ECG AT CENTRAL Monitor The ECG measured with the monitor ECG is not being sent to
the Information Center via the telemetry device.
NO ECG SOURCE Telemetry A telemetry device is paired with the monitor but the
Information Center is not detecting a valid ECG signal from
either of them.
INOP Message, Indication Source What to do