Technical Alarm Messages (INOPs) 5 Patient Alarms and INOPs
79
<SpO
2
label> NON-PULSAT.
Numeric is replaced by -?-
INOP tone.
SpO
2
Check the perfusion at measurement site. If necessary,
stimulate circulation or change measurement site. If the INOP
is due to NBP measurement on the same limb, wait until the
NBP measurement is finished.
<SpO
2
label> NO SENSOR
Numeric is replaced by -?-
INOP tone.
SpO
2
Make sure the SpO
2
sensor is connected. If the INOP persists,
try another adapter cable and sensor. If you silence this INOP,
the measurement will be switched off.
<SpO
2
LABEL> POOR SIGNAL
Label is displayed with a -?-
(questionable numeric)
SpO
2
The signal condition of the SpO
2
measurement is poor and
measurement accuracy may be compromised.
<SpO
2
LABEL> PULSE?
Numeric is replaced by -?-
INOP tone
SpO
2
The detectable pulsations of the SpO
2
signal are outside the
specified pulse rate range.
<SpO
2
LABEL> SEARCHING
Numeric unavailable
SpO
2
SpO
2
is analyzing the patient signal to derive Pulse, SpO
2
and
Perf values. Please wait until the search analysis is complete.
<SpO
2
label> SENSOR MALF
Numeric is replaced by -?-
INOP tone.
SpO
2
The SpO
2
sensor or adapter cable is faulty. Try another adapter
cable and sensor. If the INOP persists, contact your service
personnel.
<SpO
2
LABEL> SENSOR OFF
Numeric is replaced by -?-
INOP tone
SpO
2
The SpO
2
sensor is not properly applied to the patient. Apply
the sensor following the instructions supplied by the
manufacturer.
<SpO
2
LABEL> UNKN.SENSOR
Numeric is replaced by a -?-
SpO
2
The connected sensor or adapter cable is not supported by the
SpO
2
measurement. Use only specified sensors and cables.
<SpO
2
LABEL> UNPLUGGED
Numeric is replaced by -?-
INOP tone
SpO
2
An SpO
2
measurement label has been deactivated, either by
unplugging a module, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module
or reactivate the measurement label in the Measurement
Selection window.
<SpO
2
LABEL> UPGRADE
Label is displayed with a -?-, numeric is
unavailable
SpO
2
The SpO
2
measurement is currently in UPGRADE mode.
Monitoring is not possible in this mode.
Sp - vO
2
CHK SOURCES
Numeric is replaced by -?-
Sp - vO
2
Not all measurements or values required to perform the
calculation are available. Check measurement sources.
Sp - vO
2
CHK UNITS
Numeric is replaced by -?-
Sp - vO
2
The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
SRR INTERFERENCE
INOP tone
Monitor The short range radio connection has interference from
another device. Try using another channel.
SRR INVALID CHAN
INOP tone
Monitor The channel configuration of the Short Range Radio is invalid.
Check channel and channel mask configuration.
SRR MALFUNCTION Malfunction in the short range radio device. If the INOP
persists contact your service personnel.
SVR/SVRI CHK SOURCES
Numeric is replaced by -?-
SVR/SVRI Not all measurements or values required to perform the
calculation are available. Check measurement sources.
SVR/SVRI CHK UNITS
Numeric is replaced by -?-
SVR/SVRI The monitor has detected a conflict in the units used for this
calculation. Check the unit settings.
INOP Message, Indication Source What to do