Chapter 3 Calibration
Revision W HF Series X-ray Generators - Service Manual
3-14 Quantum Medical Imaging, LLC
Ion Chamber/Receptor Configuration
The Ion Chamber Configuration Screen, shown below, displays available ion
chambers that can be interfaced to the HF Series X-ray generator. This
screen also provides relay on/off settings for activating either the table and/or
wall receptor when the receptor keys on the OCP are pressed.
Note: On systems equipped with a "Standard"
type AEC (i.e., the generator is equipped with
standard type AEC Board A11, AY40-031S), the
"Ion Chamber Wall 2" and "Ion Chamber Auxil-
liary" selections are not used. These choices are
only effective if the generator is configured with
"Universal" type AEC (i.e., if generator is equipped
with universal type AEC Board A11, AY40-027S").
Figure 3-11. Ion Chamber Configuration Screen
1. Press the key located next to the ion chamber you want to configure:
Wall Ion Chamber, Table Ion Chamber, Ion Chamber Wall 2, or
Ion Chamber Auxilliary (Ion Chamber Wall 2 and Ion Chamber
Auxilliary only used on systems equipped with AEC Board A11, AY40-
027S). Be sure that the selected ion chamber is properly interfaced to
the AEC Board A11 as described in Chapter 2, ASSEMBLY & INSTALLA-
TION.)
NOT USED ON
"STANDARD"
AEC SYSTEMS
PRESS THESE
KEYS TO
SELECT ION
CHAMBER
TO
CONFIGURE