EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
Chapter 3 Calibration
Revision W HF Series X-ray Generators - Service Manual
3-14 Quantum Medical Imaging, LLC
Ion Chamber/Receptor Configuration
The Ion Chamber Configuration Screen, shown below, displays available ion
chambers that can be interfaced to the HF Series X-ray generator. This
screen also provides relay on/off settings for activating either the table and/or
wall receptor when the receptor keys on the OCP are pressed.
Note: On systems equipped with a "Standard"
type AEC (i.e., the generator is equipped with
standard type AEC Board A11, AY40-031S), the
"Ion Chamber Wall 2" and "Ion Chamber Auxil-
liary" selections are not used. These choices are
only effective if the generator is configured with
"Universal" type AEC (i.e., if generator is equipped
with universal type AEC Board A11, AY40-027S").
Figure 3-11. Ion Chamber Configuration Screen
1. Press the key located next to the ion chamber you want to configure:
Wall Ion Chamber, Table Ion Chamber, Ion Chamber Wall 2, or
Ion Chamber Auxilliary (Ion Chamber Wall 2 and Ion Chamber
Auxilliary only used on systems equipped with AEC Board A11, AY40-
027S). Be sure that the selected ion chamber is properly interfaced to
the AEC Board A11 as described in Chapter 2, ASSEMBLY & INSTALLA-
TION.)
NOT USED ON
"STANDARD"
AEC SYSTEMS
PRESS THESE
KEYS TO
SELECT ION
CHAMBER
TO
CONFIGURE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals