EasyManua.ls Logo

Quantum ODYSSEY HF Series

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Calibration
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC 3-23
Set Time Screen
The Set Time screen, shown below, displays the current time setting. To
change the time setting, proceed as follows:
Figure 3-18. Setting the Time
1. Adjust the hours or minutes setting using the appropriate keys as shown
in Figure 3-18, above, until desired time is displayed.
2. To exit the screen and return to the Start Up Screen, press the key next
to the Main field.
Film/Screen Defaults Screen
The Film/Screen Default Screen (see Figure 3-19) is used to configure the
default film/screen combination for the various image receptors used with the
generator. This setting applies to AEC and APR AEC mode exposures only. If
the selected film/screen combination has not already been calibrated, it must
be calibrated according to the AEC Calibration procedures later in this chapter
using the selected film/screen combination setting.
The generator has been pre-configured with seven different film/screen com-
bination selections designated "Wall 40"", "Wall 72"", "Table 40"", "100
Speed", "200 Speed", "800 Speed", and "Film/Screen7". If the Digital Recep-
tor setting in Service Configuration Menu is enabled, an eighth option entitled
"Digital" will be also be selectable. Note that the factory default film/screen
combination names may be changed using the film/screen combination name
edit utility as described in Service Manual.
USE THESE
KEYS TO
SET HOURS
USE THESE
KEYS TO SET
MINUTES

Table of Contents

Related product manuals