Chapter 2 Assembly & Installation
Revision W HF Series X-ray Generators - Service Manual
2-18 Quantum Medical Imaging
High Voltage Cables
It is important that these instructions be followed for the proper installation
of the High Voltage Cable terminals into the tube housing and H.V. Tank
receptacles. Use proper handling to ensure cable terminals and receptacles
are clean and free of moisture. Connect High Voltage Cables, respecting
proper anode and cathode polarity, as follows:
Figure 2-12. High Voltage Cable Connection to H.V. Tank
BEFORE INSERTING
CABLE, LOOSEN
SET SCREW
COAT WITH
INSULATING
COMPOUND
BEFORE
INSERTING
1
TIGHTEN RING NUT
ONTO RECEPTACLE
2
FULLY TIGHTEN
SET SCREW
3