Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-13
Figure 6-6. Generator System Cabling Diagram
(Internal High-Speed Starter)
12
SW1
SW2
ETHERNET RJ-45
TABLE BUCKY
WALL BUCKY
TB6
TB3
INTERFACE PCB
TB1
TB1
X-RAY GENERATOR
TB2
ROTOR
PCB *
AY40-013T
AY40-023T
SC50-015T REV. M
SINGLE PHASE OPTION
CATHODE(-)
ANODE(+)
HIGH VOLTAGE
TB1
THREE PHASE OPTION
INCOMING AC POWER
L1 L2 L3
GENERATOR GROUND
N
PHASE A
PHASE B
PHASE C
NEUTRAL
AC Power
115/230 VAC
L1
'X-RAY ON' LAMP
TB5
L2
PHASE A
PHASE B
GROUND
GROUND
TOSHIBA
X-RAY TUBE
ANODE(+)
CATHODE(-)
CATHODE(-)
ANODE(+)
AID - HSS
208-260 VAC
380-480 VAC
or
50-60 Hz
STANDARD
*Note: The Standard Rotor PCB
AY40-013T is removed when the
QG-HSS AID High Speed Starter or the
QMI-HSS is present in the system.
QG - HSS
OPTION *
TB1
OPERATOR'S CONSOLE
J4
AY51-017
WALL STAND
J2
ION CHAMBER
TB1
TB1
STANDARD
HANDGRIP
TUBESTAND
DOOR INTERLOCKS
WALL
J2
AEC PCB
TABLE
J4
TB1
TABLE
VARIAN
X-RAY TUBE
TB4
J8
LOGIC PCB
J5
AY40-006S
COLLIMATOR
WALL BUCKY
COLLIMATOR
TUBESTAND
L1
N
NEUTRAL MAY BE OMITTED
IF NOT REQUIRED BY
LOCAL ELECTRICAL CODE.
AY40-031S
TECHVISION
TB1
GREEN-PIN 2
DA10
BLACK-PIN 3
DUNLEE X-RAY TUBES
CENTER PIN
1300/1400
OUTER PIN
WHITE-PIN 1
INNER PIN
115/230VAC 15A
TECHVISION
REMOTE
OPTION
NON-INTEGRATED OPTION
CONTROL
EXPOSURE
<<<--- CONDUIT --->>>
208-240 VAC
EARTH GROUND
WHITE
SHIELD
GRN/BLK
HANDSWITCH
RED
FIELD SERVICE
LAPTOP TOOL
GND
200-240VAC
20A
L1
L2
QMI - HSS
OPTION *
AY40-036S
TB2
COMMON
MAIN
THERMAL SWITCH
THERMAL SWITCH
PHASE
TB4
TS-1 PIN
GREEN-PIN 6
ORANGE-PIN 5
TS-2 PIN
AC Power
TABLE BUCKY
ION CHAMBER
QUANTUM
QUANTUM Q-WARE
UTILITY SOFTWARE
AUXILIARY HANDGRIP
M
4
GROUND
3
COMMON
2
EXPOSE
1
PREP
4
1
L1
2
L1
12
L2/N
L1
L3
L2
L1
L1
L2
1
2
2
BLACK - GND
1
RED +24VDC
3
1
RED +24 VDC
2
BLACK - GND
1
1
2
2
2
1
3
4
5
6
7
8
BUCKY READY
+24 VDC
24V GND RTN
N/A
GROUND
115 VAC RTN
BUCKY ON
115 VAC
2
1
3
4
5
6
7
8
BUCKY READY
N/A
N/A
N/A
GROUND
115 VAC RTN
BUCKY ON
115 VAC
1
2
3
4
5
6
7
+24 VDC
24V GND RTN
N/A
EXPOSURE HOLD (COLLIMATOR READY)
N/A
N/A
EXPOSURE HOLD COMMON
1
2
3
24 VAC
24 VAC RTN
GROUND
R
S
-
2
3
2
S
E
R
I
A
L
C
O
M
M
P
O
R
T
1
2
-
P
I
N
C
I
R
C
U
L
A
R
RS-422 TECHVISION POWER/DATA
OCP POWER/DATA/CONTROL 9-PIN SUB-D
1
2
3
4
5
6
1
2
3
4
5
6
1
2
1
2
1
2
3
4
5
6
7
1
2
3
4
5
6
7
MAIN
N/A
PHASE
N/A
N/A
COMMON
BLACK
N/A
RED
N/A
N/A
WHITE
THERMAL SWITCH
THERMAL SW GND
ORANGE
GREEN
MAIN
PHASE
COMMON
BLACK
RED
WHITE
N/A
N/A
N/A
N/A
THERMAL SWITCH
THERMAL SW GND ORANGE
GREEN
MAIN
PHASE
COMMON
THERMAL SWITCH
THERMAL SW GND
BLACK
RED
WHITE
N/A
N/A
GREEN
ORANGE
1
2
3
4
5
6
7
1
2
3
6
5
MAIN
PHASE
COMMON
THERMAL SWITCH
THERMAL SW GND
2
5
3
4
1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals