EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging 2-35
(Note: Perform ion chamber balance procedure in accordance with manufac-
turer’s instructions provided with chamber.)
Typically, when the HF Series x-ray generator is shipped with an AID ICX ion
chamber, the chamber jumpers are pre-set in the following configuration:
CAUTION! The ion chamber must be electrically
isolated from the bucky to ensure proper AEC sys-
tem performance.
For systems equipped with universal type AEC (AEC Board AY40-027S):
Up to four ion chambers can be used. Connect ion chambers as follows:
Wall#1 ion chamber to A11J8
Wall#2 ion chamber to A11J12
Table ion chamber to A11J10
Auxiliary ion chamber to A11J13
Note that to connect a Siemens ion chamber AEC cable to universal AEC
Board A11 requires an interface cable assembly AY51-061 for each chamber.
This cable assembly adapts the ion chamber connector to 9-pin D-Sub type
used on the AEC Board, and provides a separate 300 VDC wire for connection
to A11TB1.
Route the ion chamber cable(s) as necessary to exit through the access open-
ing in top rear side of generator chassis, and secure as required. DO NOT
route through an opening containing a High Voltage Cable (see Figure 2-13).
Interfacing to Canon Digital Radiographic Receptor (When
Applicable)
The digital radiographic receptor is interconnected to the HF Series x-ray gen-
erator via Interface Board A9, P/N AY40-034T terminal blocks A9TB7 (Wall
Stand) and A9TB3 (Table), respectively, using 4-conductor, 18AWG cable.
Refer to the digital receptor manufacturer’s installation manual for more
detailed wiring instructions.
Table 2-5: AID ICX-153 or ICX-1153 Ion Chamber Jumper Configuration
JUMPER POSITION FUNCTION
61135A W1 1-2 & 3-4 LOW ACTIVE RESET
61135A W2, W3, W4 1-2 & 3-4 LOW ACTIVE FIELD SELECT
61135A W5, W6 1-2 & 3-4 A-B-C = 2-1-3
61135A W7 2-3 POSITIVE OUTPUT SIGNAL

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals