Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-48 Quantum Medical Imaging, LLC
Optional +28 VDC Power Supply Troubleshooting
The following steps apply only to systems equipped with the internally-
mounted +28 VDC Power Supply Option (QG-28VPS). The QG-28VPS pro-
vides a regulated +28 VDC output source to power peripheral x-ray
equipment such as tube stands, wall stands, etc. If a problem with the
power supply is suspected, proceed as follows:
1. Disconnect Power Supply QG-28VPS output wires at pins P2-1
and P2-2.
2. With the generator power turned on, measure the DC output
across Power Supply QG-28VPS pins P2-1 and P2-2. It should
measure between +26.5 and +27.5 VDC on systems intended for
use with Quantum QW-420-T/-D wall stand, and between +29.5
and +30.5 VDC on systems intended for use with RS-580 Over-
head Tube Crane. Adjust voltage output if necessary using
potentiometer on power supply.
3. If no output exists at terminals P2-1 and P2-2, verify 220 VAC
input to power supply at terminals P1-L and P1-N. If 220 VAC is
present, replace power supply.
4. If no 220 VAC at power supply input, check for open Fuse F7 on
Interface Board A9 (AY40-061T used on non-digital systems/
AY40-062T used on digital systems).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals