EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
Chapter 3 Calibration
Revision W HF Series X-ray Generators - Service Manual
3-4 Quantum Medical Imaging, LLC
ACCESSING SERVICE MODE
Calibration and configuration can only be performed by accessing the X-ray gen-
erator’s Service Mode. To gain access to the Service Mode, password entry is
required. This is to prevent unintended and/or unauthorized changes to system
settings. To access the Service Menu, proceed as follows:
1. Set the ON/OFF circuit breaker CB1 on the generator cabinet to ON position.
2. Set the power on/standby switch on the Operator Control Panel (OCP) to the
standby ( ) position, then to the on ( ) position.
3. The "Welcome Screen" appears with "Testing. Stand by." displayed in the
lower portion of screen (see Figure 3-1.).
Figure 3-1. Welcome Screen
4. Within five seconds of start up, press the upper left field select key as shown
in Figure 3-1). The Password menu displays:
Figure 3-2. Password Menu
PRESS TO
ACCESS
PASSWORD
MENU
FIELD SELECT KEYS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals