EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #448 background imageLoading...
Page #448 background image
Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC D-21
GENERATOR CALIBRATION
CAUTION! Before calibrating unit, perform warm-up
procedures recommended by tube manufacturer or
the warm-up procedure described in Chapter 3 of the
Operator’s Manual.
WARNING! The following calibration procedures
involve taking x-ray exposures. All personnel
should be behind protective barriers during expo-
sure. X-rays can be dangerous unless proper pre-
cautions and procedures are adhered to.
The generator was pre-calibrated at the factory, however, the system will need to
be re-calibrated to assure proper operation, verify compliance and set AEC optical
density (on systems equipped with AEC option). Prior to performing Generator
Calibration, the Generator Configuration program must be run first as described in
the preceding paragraphs. To gain access to the Generator Calibration utility,
password entry is required. This is to prevent unintended and/or unauthorized
changes to system settings.
1. Right mouse click on the Generator Service Monitor icon "Q" located in the
lower right side of the screen.
2. Select Generator Configuration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals