Appendix D Calibration (Using Generator Service Tools)
Revision W HF Series X-ray Generators - Service Manual
D-22 Quantum Medical Imaging, LLC
3. The "Enter Password" screen appears (see Figure D-11.).
Figure D-11. Enter Password Screen
4. In the Enter Password screen, enter the default Service Technician password
(58623) using the keypad, then press Enter. This password enables access
to the Generator Calibration program.
4. Upon validation of the password, the Load Configuration menu displays:
Figure D-12. Load Configuration Menu
The Load Configuration menu offers two choices:
• Load From File - Opens a browser window to load a previously saved
configuration file stored on the Workstation Computer (refer to Factory
Reset description in this chapter for instructions on creating a
configuration file of the generator).