EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging 2-39
Interfacing to Pausch Tomography System (When Applicable)
To interface the HF Series generator to Pausch tomo system, two lengths of
10-conductor shielded cable is required (supplied with the HF Series genera-
tor). No wire terminations are required for interconnecting the cables. Refer
to Figure 2-22 for Pausch tomo/generator interface wiring. After interfacing,
secure the tomo interface cable securely using cable ties as required.
Check that jumper JP1 is not installed on Universal AEC Board A11
(AY40-027S). If installed, remove this jumper.
IMPORTANT! Be sure to re-program the tomo
sweep times on the Pausch tomography system to
correspond with available time settings selectable
at the HF Series generator Operator Control
Panel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals