Chapter 1 Specifications
Revision W HF Series X-ray Generators - Service Manual
1-8 Quantum Medical Imaging
40 kW Single Phase (Model QG-40)
Line Voltage: 208-260 VAC (±5%), 150 Amps, 50/60 Hz
Momentary Current: 300 Amps (@208VAC)
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
40 kW Single Phase (Model QG-40-5)*
Line Voltage: 380-480 VAC (±10%), 100 Amps, 50/60 Hz
Momentary Current: 170 Amps @ 415 VAC
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
32 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz
210-250 VAC (±10%), 10 Amps, 50/60 Hz
Momentary Current: 13 Amps @ 115 VAC
7 Amps @ 230 VAC
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
40 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz
210-250 VAC (±10%), 10 Amps, 50/60 Hz
Momentary Current: 13 Amps @ 115 VAC
7 Amps @ 230 VAC
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
50 kW Single Phase, Stored Energy
Line Voltage: 105-130 (±10%), 10 Amps, 50/60 Hz
210-250 VAC (±10%), 10 Amps, 50/60 Hz
Momentary Current: 13 Amps @ 115 VAC
7 Amps @ 230 VAC
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
Note: Models QG-25-5, QG-32-5, and QG-40-5 are not UL classified.