Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging 2-55
ROTOR DRIVE BOARD A10 (AY40-013T) PERFORMANCE VERIFICATION
Note: This procedure is not required on systems equipped with high
speed starter option.
Refer to Rotor Drive Board A10 (AY40-013T) Terminal Block Location Diagram in
Chapter 6, DIAGRAMS and verify Rotor Drive Board adjustments as follows:
1. Set A10R20 to approximately mid range.
2. Connect one channel of oscilloscope to A10TP15 and second channel to
A10TP14.
3. Press and hold the PREP key for approximately 3 seconds. Verify that during
the run cycle, the amplitude of the phase and main pulses are +14 VDC to
+15 VDC and waveforms are as shown in Figure 2-26 (i.e., the peaks of
waveforms at TP15 and TP14 should be between 2.5 and 3.0 milliseconds
apart).
Figure 2-26. Rotor Board Phase and Main Pulses
4. Press and hold PREP key for approximately 3 seconds.
5. Verify the peak voltage at A10TP14 after PREP has been pressed for at least 2
seconds (rotor in boost mode) is
greater than
14 V
PEAK
. If necessary, adjust
A10R20 to achieve the correct voltage.
6. Set A10R24 to approximately midrange.
7. Press and hold PREP key for approximately 3 seconds.
2.5 - 3.0 ms
14.0 to
15.0V(peak)
A10TP15
Waveform
14.0 to
15.0V(peak)
A10TP14
Waveform

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals