EasyManua.ls Logo

Quantum ODYSSEY HF Series

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B Calibration (DiRex System)
Revision W HF Series X-ray Generators - Service Manual
B-30 Quantum Medical Imaging, LLC
kV Calibration
This procedure calibrates the generator to verify accurate system measure-
ment of peak tube potential (kVp). The generator’s ability to measure peak
tube potential accurately is verified using a pre-calibrated kVp meter (Dyna-
lyzer or equivalent) as a benchmark. Proceed as follows:
1. In the Calibration Menu, press KV button (see Figure B-16). The KV Cal-
ibration screen appears:
Figure B-19. KV Calibration Screen
WARNING! To avoid injury to personnel, ensure
that system power is removed prior to making
and removing any electrical connections.
2. Connect a kVp bleeder tank in series between X-ray tube and the genera-
tor’s H.V. Transformer (A17), or use a non-invasive kVp meter.
3. Verify 50 KVP appears in the KV Calibration Screen and "Next Measure"
appears on the top right button as shown in Figure B-19.
KV INCREASE/
DECREASE
BUTTONS

Table of Contents

Related product manuals