EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #393 background imageLoading...
Page #393 background image
Appendix B Calibration (DiRex System)
Revision W HF Series X-ray Generators - Service Manual
B-30 Quantum Medical Imaging, LLC
kV Calibration
This procedure calibrates the generator to verify accurate system measure-
ment of peak tube potential (kVp). The generator’s ability to measure peak
tube potential accurately is verified using a pre-calibrated kVp meter (Dyna-
lyzer or equivalent) as a benchmark. Proceed as follows:
1. In the Calibration Menu, press KV button (see Figure B-16). The KV Cal-
ibration screen appears:
Figure B-19. KV Calibration Screen
WARNING! To avoid injury to personnel, ensure
that system power is removed prior to making
and removing any electrical connections.
2. Connect a kVp bleeder tank in series between X-ray tube and the genera-
tor’s H.V. Transformer (A17), or use a non-invasive kVp meter.
3. Verify 50 KVP appears in the KV Calibration Screen and "Next Measure"
appears on the top right button as shown in Figure B-19.
KV INCREASE/
DECREASE
BUTTONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals