Chapter 1 Specifications
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging 1-7
NOTE: All single-phase x-ray generator input ratings assume the use of a
single-phase distribution transformer. When utilizing two legs of a 3-
phase transformer, review the National Electrical Code and any appli-
cable local electrical code(s) to ensure the correct rating distribution
transformer (in terms of KVA) is used to assure line stability. This
configuration would require an increase in the overall power (KVA) of
the distribution transformer, depending on whether the input line
transformer is Wye or Delta type. Under no circumstances should a
Delta type transformer be used for input power to a 3-phase genera-
tor.
SINGLE-PHASE X-RAY GENERATORS
Non-Stored Energy Single-phase x-ray generators require three-wire line inputs in
either of the following two configurations:
one hot wire, one neutral wire, and one earth ground wire
two hot wires and one earth ground.
Refer to Table 1-1 for power requirements of single-phase (non-stored energy)
HF Series Generators.
25 kW Single Phase (Model QG-25)
Line Voltage: 208-260 VAC (±5%), 100 Amps, 50/60 Hz
Momentary Current: 200 Amps (@208VAC)
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
25 kW Single Phase (Model QG-25-5)*
Line Voltage: 380 - 480 VAC (±10%), 60 Amps, 50/60 Hz
Momentary Current: 110 Amps @ 415 VAC
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
32 kW Single Phase (Model QG-32)
Line Voltage: 208-260 VAC (±5%), 120 Amps, 50/60 Hz
Momentary Current: 240 Amps (@208VAC)
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation
32 kW Single Phase (Model QG-32-5)*
Line Voltage: 380 - 480 VAC (±10%), 75 Amps, 50/60 Hz
Momentary Current: 140 Amps @ 415 VAC
Long Term Current: less than 1.0 Amp
Mode of Operation: Intermittent operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals