Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-4 Quantum Medical Imaging, LLC
SERVICE MAINTENANCE
WARNING! Always disconnect the equipment
from the main power supply prior to any cleaning.
WARNING! STORED ENERGY (SE) systems have
live high voltage present even with system pow-
ered off. Areas where high voltage is present
includes the battery tray fuse holders, contactor
K1, rectifier DB1, and Line Monitor/Battery
Charger Board A8.
To keep the HF Series X-ray Generator operating properly, it must be inspected
and calibrated at regular intervals, depending on the age and use of the equip-
ment. When there is a problem, call a Quantum Medical Imaging-recognized ser-
vice technician. Do not operate equipment until all repairs are complete.
Periodic planned maintenance is required to maintain the system in good working
order and to insure operational reliability. It is recommended that the Service/
Support Technician be prepared with the standard tools and equipment required
for the service call.
Visual Inspection
A complete series of inspections and functional checks was conducted at
installation to insure proper operation of the system. The following
inspection and adjustment procedures are recommended to maintain the
system in its original operating condition.
Perform Every Six Months:
Check for evidence of loose hardware or loose wires
Conduct a general inspection for worn or damaged parts
Verify that all ground conductors are properly and securely
installed and free of corrosion or damage
Check all electrical cabling and wiring for wear and fraying
Check for any indications of oil leaking from the H.V. Tank
Perform Every Twelve Months:
Verify accuracy of kVp, mA, and filament calibrations and, if
equipped, check AEC calibration (refer to Chapter 3,
CALIBRATION)
Use the Reproducibility, Linearity, and Maximum Deviation data test pro-
cedures and Calibration Data Sheets provided at the end of Chapter 3,
and verify that all calibrations are within specified tolerances as stated in
Chapter 2, SPECIFICATIONS; this is required to demonstrate compliance.
If the generator is interfaced to an external imaging system that is used
as the generator’s operator control panel, (e.g., an Agfa NX Workstation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals