Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC
5-15
Cathode
Volt
(Cont’d)
8. AC line voltage is out of specifica-
tion during an exposure. Check
the capacitor (C9-C12 and C13-
C16 on single-phase units, C1-C4
on SE units, or C9-C12 and C13-
C16 on three-phase units) voltage
is within range (i.e., 350 to 440
VDC).
9. Verify that all grounds are good
(JP2-JP9 installed or shorted on
Logic Board A1 (AY40-006S).
10. Verify that the power is sufficient
to support the exposure (i.e.,
check for weak AC Line).
Collimator Automatic collimator not respond-
ing (only appears on systems
using an automatic collimator).
1. Positive beam limitation (PBL) cir-
cuit is open; check PBL cable con-
nection between generator and
table.
2. Verify jumper JP1 is not installed
on Interface Board A9 (AY40-
023T).
Configura-
tion Load
Error {text
string}
An error occurred while loading
the configuration. Note: {text
string} will contain additional
details.
Call Technical Service/Support.
Door Entrance door to x-ray room is
open.
Verify x-ray room door is closed and
re-take exposure.
Early Term Exposure terminated prematurely
because EXPOSE key was
released before exposure was
completed.
Press any key/button on Operator
Control Panel to clear warning mes-
sage. Re-take exposure.
Error - An
error
occured
while load-
ing a view.
Please
check the
APR tree.
When trying to load a view's
parameters an error occurred.
This is typically because of an
invalid APR tree (missing value or
out-of-range value).
Load APR editor and fix
{0} = Location i.e., generator, collima-
tor, etc.
{1} = AnatomicalRegion
{2} = ViewName
{3} More details about this error.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action

Table of Contents