EasyManua.ls Logo

Quantum ODYSSEY HF Series

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-64 Quantum Medical Imaging, LLC
DiRex Integrated System - Statistics Screen
The Statistics Screen is accessed as follows:
1. Set the ON/OFF circuit breaker CB1 on the generator cabinet to
ON position.
2. Set the power on/standby switch on the Q-Connect Control Panel
(OCP) to the standby ( ) position, then to the on ( ) position.
The green power LED lamp should illuminate.
3. From the Main OCP Overlay Menu, double click the generator sta-
tus icon. The "Enter Password" screen appears.
4. In the Enter Password screen, enter the default Service Techni-
cian password (58623) using the keypad, then press Enter.
5. Upon validation of the password, the Service Options Menu dis-
plays. Press the Generator Service button.
Figure 5-5i. Service Options Menu

Table of Contents

Related product manuals