Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC
5-71
Events/Warnings Log
An event/warning log viewing program (
Log4View
) is used to capture and
view errors and other events that occur during use of the integrated X-ray
generator (Digital Radiographic System integration), as with the DiRex
System or a QG-DIG-CXDI Canon/Quantum Integration System. These
errors and events may be useful to help diagnose problems. Log4View is
a separate program that is accessible outside of the digital operations.
Log4View is installed along with the DiRex, QG-DIG-CXDI Canon/Quan-
tum integration, or NX-DR Quantum/Agfa software installation. It leaves
an icon on the desktop that will allow the log to be viewed at any time.
Every time any of the interrelated Quantum Programs that form the inte-
gration make a significant action, this is recorded in a file. This includes
normal operations as well as warnings and errors because all may be use-
ful in both viewing normal system operation, as well as abnormal (error)
conditions.
Refer to Appendix C, LOG4VIEW USER INSTRUCTIONS for details about
using the Log4View application.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals