Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-48
Figure 6-41. QMI High-Speed Starter Circuit Block
Diagram
J3
TB4
K1
TP3
-15V
TP7
-15V
TP4
+5V
J3
D27
MBRS360TR
TP2
I feedback 1
TP19
ROTATING
TP18
18VDC
TP14
V feedback
J4
K3
TQ2SA-12V
P5
TB3
J1
R59
100 2W
TP4
RUN ON
J2
P1
P2
J5
TP16
60/180 Hz
TP1
GND
P5
TP11
OVERCURRENT
TP15
I feedback 2
TP5
DRIVE 2
TP8
HIGH/LOW SPEED
TP6
DRIVE 1
F4
FUSE FLQ15
D29
SMCG18
D4
1N4005
TP2
+15V
TB2
PD7
F6 FUSE THERMAL 0.5A
P5
TB1
CON7
J1
TP9
15V
TP6
I feedback 2
C6
100uf 25V
+
C1
100uf 25V
+
TP7
BOOST ON
F5
FUSE 0.1A 250V
P3
F5
FUSE 0.1A 250V
TP10
GND
J2
CON9
PD6
F5
FUSE 15A 500V
P4
-15V
+15V
5V
-15V
+15V
-15V
+15V
+15V
+15V
TUBE_TEMP
BOOST
RUN
ROTOR ON
SPEED HIGH/LOW
1
2
3
4
5
6
7
8
1
2
4
3
A
B
1
1
1
1
2
3
1
1
1
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
3
2
4
8
9
7
1
10
1
2
3
4
5
6
1
2
3
4
5
1
2
3
4
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
1
2
3
4
1
2
3
4
5
6
7
8
9
10
11
12
13
14
1
2
3
4
5
6
1
1
1
2
3
4
5
6
1
1
1
1
1
1
2
1
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
7
1
2
3
4
5
6
7
8
9
10
1
1
1
1
2
3
4
5
6
7
8
1
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
11
12
13
14
BOOST
RUN
CURR_LIMIT
C_FEEDBACK
RESET
V_FEEDBACK
SPEED
ROTOR ON
ROTOR ON
RTN
380VAC
80VAC
220VAC
60VAC
36VDC
RTNRESET
HIGH/LOW SPEED
TEMP. FAULT
ROTOR ON
FAU LT
BOOST ON
ROTOR ROTATING
RUN ON
RTN
AY40-035S
VOLTAGE
CONTROL BOARD
REGULATOR
ISOLATION &
LEVEL
CONVERSION
SYNCRONIZING
CIRCUITS
60 Hz/180 Hz
GENERTOR
DRIVE CONTROL
CIRCUIT
DRIVER
DRIVE
FREQUENCY
MONITOR
FROM LOGIC BOARD
CONTROL
TIMING CONTROL
& LIMITS
ANALOG
CONTROL
CIRCUITS
RELAYS
CONTROL
CIRCUITS
AY40-036S DRIVE BOARD
FROM T2 INPUT VOLTAGE
RELAYS
CONTROL
BRIDGE RECTIFIER
AND FILTER
IGBT H BRIDGE AND
GATE DRIVERS
MAIN
12uF
12uF
COMM
PHASE
QUAD OUTPUT 14VDC
POWER SUPPLY
25uF
CURRENT SENSING
& FUALT PROTECTION
VOLTAGE SENSING
TEMP.
TUBE
DEPENDENT
FROM POWER SUPPLY
TO TUBE
ROTOR
TO TUBE
TEMP.
SWITCH
FROM POWER
SUPPLY
OUTPUT 1
OUTPUT 2
OUTPUT 3
OUTPUT 4
DRIVE 1
DRIVE 2
SC50-069 REV. A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals