Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-53
Figure 6-46. Line Monitor Board A8 (AY40-022T1),
Single-Phase Systems - Test Point/Jumper Location
Diagram
FUSES F1, F3
FUSE F2
3 AMPS
USED TO TURN ON GENERATOR
AND FOR EXTERNAL TUBE STAND POWER SUPPLY)
+30 VDC OUTPUT
AC INPUT PROTECTION
AC INPUT PROTECTION
3A
1/4A (FOR 480 VAC INPUT)
1/2A (FOR 240 VAC INPUT)
1/4A (FOR 480 VAC INPUT)
1/2A (FOR 240 VAC INPUT)
FUSE F2
FUSE F1
FUSE RATING FUNCTION
FUSE F3
43
0 VA
C
3
8
0
V
A
C
2
4
0
V
A
C
2
4
0
V
A
C
2
4
0
V
A
C
4
8
0
V
A
C
JUMPER
JUMPER
4
3
0
V
A
C
3
8
0
V
A
C
2
4
0
V
A
C
2
4
0
V
A
C
240 VAC
480 VAC
AC INPUT VOLTAGE
JUMPER POSITIONS:
200-260 VAC, 1PH
LINE INPUT
INSTALLED
JUMPER
240 (3 PLACES
SHOWN ABOVE)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals