Chapter 6 Diagrams
HF Series X-ray Generator - Service Manual Revision W
Quantum Medical Imaging, LLC 6-53
Figure 6-46. Line Monitor Board A8 (AY40-022T1),
Single-Phase Systems - Test Point/Jumper Location
Diagram
FUSES F1, F3
FUSE F2
3 AMPS
USED TO TURN ON GENERATOR
AND FOR EXTERNAL TUBE STAND POWER SUPPLY)
+30 VDC OUTPUT
AC INPUT PROTECTION
AC INPUT PROTECTION
3A
1/4A (FOR 480 VAC INPUT)
1/2A (FOR 240 VAC INPUT)
1/4A (FOR 480 VAC INPUT)
1/2A (FOR 240 VAC INPUT)
FUSE F2
FUSE F1
FUSE RATING FUNCTION
FUSE F3
43
0 VA
C
3
8
0
V
A
C
2
4
0
V
A
C
2
4
0
V
A
C
2
4
0
V
A
C
4
8
0
V
A
C
JUMPER
JUMPER
4
3
0
V
A
C
3
8
0
V
A
C
2
4
0
V
A
C
2
4
0
V
A
C
240 VAC
480 VAC
AC INPUT VOLTAGE
JUMPER POSITIONS:
200-260 VAC, 1PH
LINE INPUT
INSTALLED
JUMPER
240 (3 PLACES
SHOWN ABOVE)