EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #411 background imageLoading...
Page #411 background image
Appendix B Calibration (DiRex System)
Revision W HF Series X-ray Generators - Service Manual
B-48 Quantum Medical Imaging, LLC
1. From the Calibration Menu (see Figure 3-16), press the AEC button. The
Select Film Screen Menu is displayed:
Figure B-25. Select Film Screen Menu
2. Select the film/screen combination to calibrate (use the Next button to
display additional choices).
The HF Series X-ray generator provides for calibration of up to nine different
film/screen combinations. Note that each image receptor used with the gen-
erator is configurable for use with a particular film/screen combination using
the Film Screen Default Settings Screen described previously in the Configu-
ration section of this chapter (see Figure 3-13). For example, when the oper-
ator presses the WALL button on the operator control panel and selects the
WALL 72" receptor, the system can be programmed to use the "Wall 72"" des-
ignated film/screen combination (i.e., the film/screen type used with the wall
stand at 72" SID). On the same system, a different film/screen type, desig-
nated as "Table 40" for example, could be configured for use with the TABLE
receptor.
Typically, on a system using a wall stand and radiographic table, three sepa-
rate film/screen combinations are used:
Wall 72" - Used when the operator selects wall receptor at 72 inch
source-to-image distance (SID)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals