Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC D-9
3. The "Enter Password" screen appears (see Figure D-2.).
Figure D-2. Enter Password Screen
4. In the Enter Password screen, enter the default Service Technician pass-
word (58623) using the keypad, then press Enter. This password
enables access to the Generator Configuration program.
5. Upon validation of the password, the Load Configuration menu displays:
Figure D-3. Load Configuration Menu
The Load Configuration menu offers two choices:
• Load From File - Opens a browser window to load a previously saved
configuration file stored on the Workstation Computer (refer to Factory
Reset description in this chapter for instructions on creating a
configuration file of the generator).