EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #485 background imageLoading...
Page #485 background image
Appendix D Calibration (Using Generator Service Tools)
Revision W HF Series X-ray Generators - Service Manual
D-58 Quantum Medical Imaging, LLC
6. If density value is too low, decrease the Offset value displayed on the
Base OD Calibration screen (Step 2.) by adjusting the slide control, or by
selecting the data entry box and typing the value using a keyboard. The
Offset value adjustment range is from 500 to 4000. Increase or decrease
the Offset value as required, then press Calibrate.
7. Press Repeat button to re-take exposure.
8. Continue until the measured density value equals the Target Level optical
density setting to within DR System manufacturer’s specification. When
it does, press Confirm button. The system advances to AEC kVp Calibra-
tion.
AEC kVp Calibration Procedure
The AEC KV Calibration screen (shown below) is used to calibrate the tube
voltage to the Base (Target Level) optical density setting. There are a total of
six kV calibration points for a 125 kV system (50, 60, 80, 100, 110, and 120
kV); 150 kV systems require an additional two kV calibration points (130 and
140 kV). The image receptor and chamber field settings remain the same as
those used in the Base OD Calibration procedure described previously.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals